T: 0114 275 0387 sales@plattsnisbett.com

Instructions for Reprocessing of Reusable Devices

The following instructions are for all reusable medical devices supplied by Platts & Nisbett Ltd, unless stated otherwise with the packaging of the product. These instructions are intended for use only by persons with the required specialist knowledge and training.

Warnings

Note: When reprocessing medical devices, always handle with care, wearing protective clothing, gloves and eyewear in accordance with local Health & Safety procedures.

Limitations on reprocessing

Instructions

From point of use

Preparation for decontamiation

Cleaning: Automated

Note: Automated cleaning may not be suitable for all lumens and cannula in which case clean manually with a water jet gun, if available, and an appropriate brush (and stilette if provided) that reaches the depth of the feature. After manually cleaning, pass all devices through an automatic cleaning cycle to achieve disinfection.

Note: These instruments have been validated using a washer-disinfector cycle validated to include two cold rinses at <35ºC, a detergent cycle and a rinse cycle both at >50ºC, a disinfection cycle operating at a temperature of between 80ºC and 90ºC for a minimum holding time of 1 minute (actual holding time in excess of 2 minutes 50 seconds) and a 20 minute drying cycle. The detergent used was low foaming, non-ionising spray wash detergent cleaner (max 12pH) and the rinse aid a neutral pH low foaming, non-ionic surfactant with isopropyl alcohol.

Cleaning: Manual

Note: Manual cleaning is NOT a disinfection process: When manual cleaning is used it may not be possible to disinfect the device prior to further handling.

Cleaning Inspection

Maintenance

Inspection and functional testing

Note: If an instrument is returned to the manufacturer / supplier, the instrument must be decontaminated and sterilised and be accompanied with the relevant documented evidence.

Packaging

Sterilisation (USA)

Sterilisation (Outside USA)

Storage

Additional Information

Note: It is the responsibility of the reprocessor to ensure that the reprocessing as actually performed using equipment, materials and personnel in the reprocessing facility achieve the desired result. This requires validation and routine monitoring of the process. Likewise any deviation by the reprocessor from the instructions provided must be properly evaluated for effectiveness and potential adverse consequences.

Download these instructions

© Platts & Nisbett Limited 2013 | Site Map | Blog Website managed by Pause